Инструменты пользователя

Инструменты сайта


error_19_bucket_full_shaker_touch_2_en

Error 19 BUCKET FULL

Reasons for this error:

  • Waste container is full
  • Waste sensor float is positioned incorrectly

The error appears when the diode on the sensor is lit. Empty the container and ensure that the waste sensor is fixed at container. To reset the error, press CONTROLLER REBOOT. Make sure that the protective film is removed on the sensor part.

error_19_bucket_full_shaker_touch_2_en.txt · Последнее изменение: 2026/03/18 14:39 — admin